? E115l Several relays not set to the same failure mode are assignedto the pump controlà All relays which are assigned to the pump control must be setto the same failure mode? E116l Several relays are assignef to the pump control which are notconfigured to the same modeà All relays which are assigned to the pump control must be setto the same modeDepending on the failure reason and measures taken, the stepsdescribed in chapter "Set up" must be carried out again, if necessary.9.3 Instrument repairIf a repair is necessary, please proceed as follows:You can download a return form (23 KB) from our Internet homepagewww.vega.com under: "Downloads - Forms and certificates - Repairform".By doing this you help us carry out the repair quickly and withouthaving to call back for needed information.l Print and fill out one form per instrumentl Clean the instrument and pack it damage-proofl Attach the completed form and, if need be, also a safety datasheet outside on the packagingl Please ask the agency serving you for the address of your returnshipment. You can find the respective agency on our websitewww.vega.com under: "Company - VEGA worldwide"Reaction after fault rec-tification46 VEGAMET 624 • 4 … 20 mA/HART signal conditioning instrument9 Maintenance and fault rectification28969-EN-091116