1IntroductionThe M Series CCT is designed with several user selectable features that allow operators toconfigure the device according to their protocols and local requirements. This manual describesthe configuration options and how to change them. This manual also lists the configurablefeatures, options and default settings. Some features are available to manual mode operatorsonly.The default settings shown below are for North American units only. To print the currentconfiguration of your unit, while the unit is in the configuration mode, press the Print Configsoftkey.Configuring the UnitAfter the unit has been off for 10 seconds or more, simultaneously press and hold the rightmostand leftmost softkeys. Turn the selector switch to MONITOR or ON while holding the buttonsdepressed for approximately four (4) seconds. After four (4) seconds the following screen willappear.Users entering System Configuration for the first time must enter the default passcode(00000000) for System Configuration to continue.Press the 0 (zero) softkey. This moves the highlight to the next digit. Repeatedly press the 0(zero) softkey until all eight digits have been entered. The Configuration Passcode screen willthen automatically advance to the Select Language menu.To change one of the entered digits use the Delete softkey to highlight the incorrect digit andreenter the correct code.CAUTIONThe following features are not currently implemented on the M Series CCT: Voice Markers andAuto Self-Test.6\VWHP&RQILJXUDWLRQ 'HOHWH(QWHU&RQILJXUDWLRQ3DVVFRGHRU3RZHU2IIWR([LW""""""""