SET-UPAxio Zoom.V16 Installing and connecting the human interface panel (HIP) Carl Zeiss05/2012 435080-9030-701 233.5 Installing and connecting the human interface panel (HIP)The HIP can be swiveled by about 30° around its longitudinal axis to achieve a convenientviewing angle. Be sure not to exceed the stated swivel range. Forcible twisting may result indamage to the device.The plug-in power supply is supplied with four plug adapters (EURO, US, UK, AUS). Beforeconnecting it to the power outlet, choose the appropriate adapter and push it onto the plug-inpower supply.Install the HIP (Fig. 8/3) to the microscope body (zoom function) or motorized focusing drive (focusfunction) as follows:Fig. 8 Installing the HIP control unit• Remove the cover (right) from the microscope body or focusing drive (Fig. 8/1).• Pull out the connecting cable (Fig. 8/7) slightly without applying force and connect it carefully to thecorresponding HIP connector socket.• Push the cable fully back into the casing without bending it; carefully attach the HIP so that the nose(Fig. 8/2) on the HIP engages in the groove provided (Fig. 8/8).• Screw in the socket head screw (SW 3) (Fig. 8/6) as far as it will go using the ball-head screwdriver(Fig. 8/5).• Push the cover cap onto the HIP (Fig. 8/4).• Connect the plug-in power supply to the mains supply.