000000-2121-813-KurzGA01-GB-280815WARNING – GENERAL HAZARDSThese quick instructions are intended as an overview. Instructions for the safe operation of the device are to be found in the user's manual.F Please read the CIRRUS photo user manual carefully before using this product.Select acquisition mode and scan parametersSelect this mode forOCT imaging onlySelect this modefor combined OCT+ fundus imagingSelect this modefor fundus imagingonly.Select the desired imaging mode.Advanced tip:For patients who have difficulty fixating, take OCT and fundus images asseparate acquisitions (CIRRUS and Photo), and combine them usingor .Align and fo cus the instrument using working distance dotsPull the instrument back from the patient and place patient‘s chin in the chinrest.Adjust table height and use eye level marks to adjust height of chinrest.Use joystick to align instrument with patient pupil (use alignment circles for bestpositioning).Move instrument towards patient. The focusing aid and working distance dots (twowhite dots) will appear together with the fundus image.Bring the working distance dots roughly into focus.If AutoFocus fails, use knob on side of instrument (bars should be alignedhorizontally).Bring the working distance dots into focus by turning and moving the joystick andalign the working distance dots within the horizontal lines.Acquire image(s)Instruct patient to blink once and to look directly at the fixation target .Click joystick button to capture image(s).Advanced tip:To modify OCT scan pattern location, press arrow () on keypad before capturingimages. Use controls on OCT fundus image to change the scan location androtation/size (HD Raster only).Grab and drag scanpattern to preferred scanlocation.Grab edge of scan pattern toresize the raster scan lengthand spacing. For 1 Line Raster,drag spacing to zero.Press on keypad or theOptimize button in display.Use the scroll wheel of yourmouse to center the scan.Review the captured scan, then save or try again.Evaluate the B-scan quality and the signal strength (should be 6 or greater).Check that the OCT B-scans in the display windows appear centered and that nodata is cut off or missing.Click Try Again to repeat scan acquisition, if necessary.Click Save to save the OCT scan (fundus images are automatically saved).false correctCIRRUS ™ photo Scan Acquisition (Models 600 and 800)Quick Reference GuideCMYCMMYCYCMYK00542 BBP CIRRUS photo 600 quick guide.pdf 1 18/01/2016 22:13