18 HiToP® 184 gbo Medizintechnik AG Version 2.02.4.4 Modification of the basic settingsIndication Menu selected in Unit settings:1. Press the black Menu key.2. You are in the display window Main menu.3. Move the scroll bar with the Modification knob to Basic settings.4. Confirm by pressing the white Softkey [Continue].You are in the menu Basic settings. The menu provides the following options:Basic settings Setting possibilities DeliveryScan – Starting frequency= lower cut-off frequency for SimulFAM® i4096 - 16384 Hz 4096 HzScan – Ending frequency= upper cut-off frequency for SimulFAM® i8192 - 32768 Hz 32768 HzTreatment time FAM X= basic setting of the therapy time which is preset uponswitching-on the device or selecting a channel.5 – 90 min 60 minTreatment time FAM i= basic setting of the therapy time which is preset uponswitching-on the device or selecting a channel.5 – 90 min 60 minSlope FAM X= basic setting of slope for SimulFAM® X which is presetupon switching-on the device or selecting a channel.10 – 200 % 150 %Blocking frequency= basic setting of the frequency for nerve block which is presetupon selecting the blocking from the Indication menu.4096 – 12274 Hz 12274 HzIntensity monitoring= see chapter 2.5 Intensity monitoringon / off onTable 4: Basic settings (Indication Menu)5. Move the scroll bar with the Modification knob to the basic setting to be changed.6. Confirm by pressing the white Softkey [Continue].7. Move the cursor or change the value with the Modification knob to the desired value.8. Confirm by pressing the white Softkey [Store].9. Exit the device setting menu by pressing the black Menu key [Return].10. On the display the Main menu appears again.11. Exit the Main menu by pressing the black Menu key [Return].12. On the display the Overview window appears again.