42 HiToP® 184 gbo Medizintechnik AG Version 2.08. If the data cannot be transmitted, the following note will be shown:MenüFIXIIIIIIIVPatient unitProgram RepeatCancelPlease place the patient unit on theside at a distance of 30 - 50 cm,change it to the programming modusand press the Softkey.Figure 27: Menu Programming of patient unit / Transmission error!9. Correct the location of the patient device and press the Softkey [Repeat].10. Upon correct data transmission the following note will be shown on the display:Successful data transmission(see Figure 26: Menu Programming of patient unit / Successful data transmission).5.2.2 Selection of channel I, channel II, channel III , channel IVHere you select the type of current and its parameters set in the respective channel. These can beSimulFAM® i, SimulFAM® X or the nerve block. These parameters can be changed in a changing maskbefore the programming.1. Move with the Modification knob to the desired channel.2. The type font will be shown inversely.3. Press the Softkey [Select].4. You move to the display of the preset type of current of the respective channel.5.2.3 Selection SimulFAM® X1. Move with the Modification knob to SimulFAM® X.2. The selection is shown inversely.3. Press the Softkey [Select].