32HiToP 1touch gbo Medizintechnik AG Version 1.29 Technical DataMains voltage andfrequency:100 – 240 V, 50-60 HzPower consumption: max. 45 VAOutput current: max. 310 mAOutput voltage: max. 76 VOutput power: max. 5000 mWPatient resistance: 30 – 1500 ΩApplied part Contact surfaces at the electrodes.Mode of operation: continuous operationMDD device class: IIaProtection degree: I acc. to IEC 60601-1Protection class: BF acc. to IEC 60601-1Protection againstingress of: liquidsIP X0Dimensions: max. 27 cm x 24,5 cm x 6 cm (W x H x D)Weight: 2,44 kg without accessoriesColor: aluminium natural anodized and grey RAL 7016Display: 10,1“ TFT LCD touch screenEnvironmentalconditions:operation of thedevice:temperature range +10 °C ... +35 °Crelative humidity of air 30 ... 75 %transport andstorage:temperature range +5 °C ... +50 °Crelative humidity < 90 %, noncondensingCurrent types:carrier frequency 4096 –32768 Hz Sinus wavelow frequency 0.1 Hz – 200 Hzaccuracy ± 10 %By request of technical personnel gbo Medizintechnik AG can offer spare part lists and circuit diagrams.The appliance inlet is used for all pin disconnection from the mains.gbo Medizintechnik AG reserves the right to modify design and specifications withoutprior notice.