HiToP 1touch 7 gbo Medizintechnik AG Version 1.21.5 Short operating instructions for SimulFAM® X(1) Connect the device with the mains switch at the back of the device. The devicewill then carry out an automatic check of all functions. If faultless, the deviceautomatically shows the home screen.(2) Connect the accessories with the respective socket on the back side of thedevice.(3) In the display you see the window with the settings of the last treatment.(4) Select SimulFAM® X from the list of “More Therapies” or from the favoriteslist.(5) Apply the electrodes to the patient (see chapter 5.3, Position of the Electrodes).(6) Increase the current slowly with the Intensity regulator until the patient feels atingling sensation.(7) In case of a frequency scan the current stimulation frequency is shown. Thefrequency values increase and decrease every second.(8) A special tone marks the end of the treatment time.(9) The current to the patient will decrease automatically.2 Start of Operation2.1 Transport and AssemblyAfter unpacking check all components (unit with power cable, electrode connectioncables and electrodes) for external integrity.The unit may be placed on any flat surface. It must not be placed on the floor. Keep awall distance of at least 20 cm. The device should neither be placed in front ofradiators nor should it be covered by pillows or blankets while in operation. Thedevice is not made for outdoor operation. Do not operate the device inside thetransport bag.2.1.1 AssemblingOn delivery, the two feet for placing the appliance are kept into the correspondingholes. The feet are held by magnets. Remove the two feet and screw them clockwiseinto the holes provided on the bottom of the device.