24 N°: 999341_EN_DVR 2pro/12.07.2017Assembly and connectionConnection via small flangeRequired connection material: Clamping ring with centering orcentering ring for KF DN16 (tool: fork wrenche size 17).1. Place the small flange KF DN16 on top of vacuum connectionof the gauge.2. Screw in the small flange KF DN16 hand tight.3. Remove the protection cap (g) from the small flangeKF DN16 (h).4. Put the gauge with the centering onto the connection of theapparatus small flange KF DN16.5. Fix the vacuum gauge with a clamping ring (i).i(g)H