Micro-Ohmmeter Model 6250 393. “- - - -” if the temperature sensor is activated but isincorrectly connected or not connected at all4. The measured temperature is out of range (-10° to 55°C)If the temperature is out of range or if the sensor leads are disconnected,the Model 6250 displays “Err 10”.Temperature compensation can be toggled ON or OFF after the mea-surement is completed, in resistive and inductive modes or at any time inAUTO mode.4.7 Activating AlarmsThe alarm programming menu offers the option of choosing one or twoalarm thresholds. An alarm comprises a value, an activation direction(above ▲ or below ▼ the setpoint) and a sound level setting for the buzzer,should the alarm become active.NOTE: Alarms are activated by successively pressing the button fol-lowed by the ALARM button.The display will indicate one of the following after each set of buttonpresses:• Alarm 1 and the activation direction, ▲ or ▼• Alarm 2 and the activation direction, ▲ or ▼• Alarm 1 and Alarm 2 and activation directions, ▲ or ▼ for eachThese values and direction parameters are programmable. See Settingan Alarm Set Point, Direction and Buzzer Levels (§4.2.12). The buzzer willsound when Alarm 1 and/or Alarm 2 threshold values are reached.www. .com information@itm.com1.800.561.8187