Clamp-on Ground Resistance Tester Model 6417 316.1.2.2 Measurement ResultsFirst Screen:Once the measurement has stabilized, thedisplay shows the first screen, which indicates:- The leakage current.- The loop impedance referenced to thechosen frequency.The impedance is measured only if the leakagecurrent is less than 10A. In the 10 to 40A range,only the current is displayed; the NOISE symbolblinks and the impedance is replaced by dashes.Figure 6-2Second Screen:Press ► to display the second screen, whichindicates the contact voltage (product ZxI).Figure 6-3Third Screen:Press ► to display the third screen, whichindicates the values of R and L. The loopinductance and loop resistance are displayed.When the inductive component is negligible* withrespect to the resistive component, the symbolR=Z is displayed, and only the impedance isdisplayed; the inductance is replaced by dashes.* R > 25W or R[W] / L[H] > 105Figure 6-46.1.3 General InformationThis information is displayed in both Standard and Advanced modes.PRODUCT ZxI GREATER THAN 50V- The blinking NOISE symbol is displayed.- The impedance blinks.- The hazardous voltage symbol blinks.Figure 6-5www. GlobalTestSupply .comFind Quality Products Online at: sales@GlobalTestSupply.com