LifeSource Health Line (toll-Free): 1-888-726-9966E-2Display Symbol Condition/Cause recommended actionThe symbol shown while mea-surement is in progress. It blinkswhile detecting the pulse.Measurement is in progress,remain as still as possible.An irregular heartbeat or bodymovement may have occurred.Refer to page E-8 for more infor-mation on irregular heartbeats.Take measurement again andconsult with your physician.Previous measurement stored inMEMORY.Battery FullThe battery power indicator dur-ing measurement.Replace all batteries with newones when the symbol blinks.Alkaline batteries recom-mended.Battery LowThe batteries are low when itblinks.Unstable blood pressure due toexcessive body movement.Try the measurement again.Remain very still during themeasurement.The systolic and diastolic valuesare within 10 mmHg of eachother.Fasten the cuff correctly, and trythe measurement again.The pressure value did notincrease during inflation.Check for air leaks along thetube and around the air socket.The cuff is not fastened correctly. Refasten the cuff and retakemeasurement.There is an air leak in the cuff ormonitor.Make sure tube is properly con-nected to cuff and monitor.The pulse is not detected correctly.The symbol will appear on the pulsedisplay area.Try the measurement again.Remain very still during themeasurement.Cuff Inflation Meter Measurement is in progress,remain as still as possiblePressure Rating Indicator™ Refer to the section “AboutPressure Rating Indicator” forfurther explanation.WhaT disPlay syMBols MeanUA-631V Manual_English.indd 2 8/5/09 8:08:39 PM