18 | USAFINION™ 2 User ManualConnectivity settingsSee Table 2 for description of the available Connectivity settings.Touch in General settings to enter Connectivity settingsSelect appropriate communication protocolASTM HLASTM LLHL7POCT1-ACommunication protocol is disabled as default.Select which results to be transferred to LIS/HIS/EMR by selecting the appropriate buttonPatient results onlyPatient and control resultsSelect “New results only” and previous obtained results will not be transferred to the LIS/HIS/EMRTouch to enter page 2 of the configuration or to return to the General settings menuTouch to enter Server IP and Port number, Receiving application (available for ASTM HL, ASTM LL and HL7 only) and Receiving facility (availablefor HL7 only).Enter the server IP address: Press to continue to the Port number setting.Enter the server port number setting: Touch to continue to the Receiving application setting or to close the text input.Enter the Receiving application setting: Press to continue to the Receiving facility setting or to close the text input.Use the button to select Patient ID as (available for HL7 only):HIS Patient IDVisit NumberTouch to return to the General Settings menuTable 2 Connectivity settingsConsult your network administrator and LIS/HIS/EMR administrator for required connectivity settings.Protocol ASTM HL ASTM High Level: The communication protocol is based on ASTM E 1394 - 97ASTM LL ASTM Low Level: The communication protocol is based on ASTM E 1381 - 95HL7 HL7: The communication protocol is based on HL7 version 2.4POCT1-A POCT1-A: The communication protocol is based on CLSI: POCT01-A2 Point-of-Care Connectivity; ApprovedStandard – Second EditionDisabled Data connectivity is disabledResults Patient results only Only patient results will be transferred to the LIS/HIS/EMRPatient and qualitycontrolBoth patient and QC results will be transferred to the LIS/HIS/EMRServer IP Insert the IP address of the receiving system [0-255.0-255.0-255.1-254]Port [0-65535] (0 = not set)ReceivingApplication(ASTM HL, ASTM LL and HL7 only)Receiving application name (0 – 30 characters)Receiving Facility (HL7 only)Receiving facility name (0-30 characters)Patient ID As HIS Patient ID (HL7 only)Visit Number (HL7 only)For further information about the connectivity settings, see the Afinion 2 data sheets for POCT1-A, ASTM and HL7 which can be obtained at www.Alere.com orthrough your local Afinion supplier.Configuration