40P8.0: Resetting to factory default settingsProgram 8.0 lets you reset all parameters to factory default settings. This clears allcalibration data, memory, and any other setup functions you might have changed.It does not clear clock settings.AB7.9From measurement mode1. Press Setup key to enter Set Up mode.2. Press the s and t keys to scroll throughsubgroups until you view parameterP8.0.See figure3. Press the ENTER key to enterparameter P8.0.See figure4. Press the s and t keys to togglebetween NO and YES.• NO retains current settings• YES resets to factory default settings5. Press the ENTER key to confirm selec-tion and to return to the subgroup menu.Press the CAL/MEAS key to return tomeasurement mode.NotesSee Addendum 5 on page 49 for a table offactory default settings.AB