4916. Addendum 5: Meter factory default settingsType Parameter Default RemarksP1.0 Memory clear no Retains current memoryP2.1 Viewing previous – – – No calibration data for 1st rangeP2.2 calibration data – – – No calibration data for 2nd rangeP2.3 – – – No calibration data for 3rd rangeP2.4 – – – No calibration data for 4th rangeP2.5 – – – No calibration data for 5th rangeP3.1 Viewing 1.0 No offset for effective cell constant (1st range)P3.2 probe data 1.0 No offset for effective cell constant (2nd range)P3.3 1.0 No offset for effective cell constant (3rd range)P3.4 1.0 No offset for effective cell constant (4th range)P3.5 1.0 No offset for effective cell constant (5th range)P4.1 Ready indicator/ READY/ON Ready indicator on; auto endpoint offauto endpointP4.2 Select °C/°F °C Degrees CelsiusP4.3 ATC on or off ATC on —P4.4 TDS factor 0.5 Adjustable from 0.4 to 1.0P5.1 Temp. coefficient 2.1% per °C Adjustable from 0 to 10%P5.2 Normalization 25°C Adjustable from 15 to 30°CtemperatureP6.1 Cell constant 1.0 Select from K = 1.0, 0.1 or 10P8.0 Factory default no Retains your current settingsNotesThe P7.0 clock function retains the year and time that you set, even after reverting tofactory default settings.