78OUR Setup: OUR ConfigurationXL40 and 60 metersThis screen allows you configure the OUR dilution factor, minimum time in minutes for the test,maximum time in minutes for the test, minimum starting DO and minimum Ending DO. All valuesfor these parameters are entered via a numeric keypad.Dilution factor can be set from 1 to 25, if no dilution was required use a factor of 1. Minimum timeshould be less than the maximum time for the test (10 character limit for time in minutes).Minimum starting DO can be set from 5 to 60. Minimum ending DO can be set from 0 to 60.To set OUR Configuration1. Touch the Touch here to edit button next to the OUR Configuration to access the OURConfiguration screen.2. Touch the Dilution Factor (1 to X) box and use the numeric keypad on the screen to enterdesired value. The BS key allows you to remove characters that are incorrectly entered.3. Touch Enter on the keypad to accept the value and return to the OUR Configurationscreen.4. Repeat steps 2 and 3 for the remaining boxes, Minimum Time (min), Maximum Time (min),Minimum Starting D.O. and Minimum Ending D.O.5. Touch the OK button to accept the changes of the entire OUR configuration and return tothe OUR Setup screen.