6-7SAFETY ANALYZER AUTOLINE VOLTAGE CHECKLine VoltageTo ensure normal operating conditions for the DUT, theLine Voltage is measured in this step. The display willindicate if Line Voltage Passed.(See ALARMS for details if it does not pass).NOTE:CThe Line Voltage must be within+20% to –10% of its rated value.STEP 4 9STEP 5 9SAFETYSAFETYSAFETYSAFETYSAFETY ANANANANANALALALALALYZER--AYZER--AYZER--AYZER--AYZER--AUTUTUTUTUTOMAOMAOMAOMAOMATIC MODETIC MODETIC MODETIC MODETIC MODERUNNING AN AUTOMATIC TESTSAFETY ANALYZER AUTOGND RESIST INVALIDLIMIT 2.00ΩΩΩΩΩMEASUREMENT --.--ΩΩΩΩΩGround Lead VerificationTo help confirm the ground lead is properly zeroed andconnected, the ground resistance measurement isverified in this step. The display will indicate if GroundResistance is invalid and will pause the test until theneeded correction is made.NOTES:CGround Resistance would be invalidif:-the lead is not properly attached tothe DUT-the lead is still attached to stud-the test lead is bad or brokenC If the lead is adjusted and themeasurement becomes valid, the testwill continue.