6-15DATALOG SAMPLE FOR AN AUTOMATIC TESTDATALOG SAMPLE FOR AN AUTOMATIC TESTSAFETYSAFETYSAFETYSAFETYSAFETY ANANANANANALALALALALYZER--AYZER--AYZER--AYZER--AYZER--AUTUTUTUTUTOMAOMAOMAOMAOMATIC MODETIC MODETIC MODETIC MODETIC MODENOTES:CThis sample shows both summaryand data information.CThere are three options for the formof the report: Summary, Data Pointsor Summary and Data.CThis option can be selected from theDatalog screen.C(See DATALOG--SAFETYANALYZER for details.)PAGE 01BC GROUPIPA-2000 INFUSION PUMP ANALYZERPUMP INFORMATION:MANUFACTURER: PUMP MAKER MODEL: 123SERIAL NUMBER: 456789DEPARTMENT: ER LOCATION: HOSPITALTEST STARTED AT: 14:59:01 ON 6/27/02SAFETY ANALYZER DATALINE VOLTAGE 125.3VGROUND RESISTANCE .08ohmLOAD CURRENT .06ANORMAL GROUND CURRENT 15uAMEASUREMENT METHODDUT: ON POL: FWDNEU:CLOSE HOT:CLOSEFAULT GROUND CURRENT 25uAMEASUREMENT METHODDUT: ON POL: REVNEU: OPEN HOT:CLOSENORMAL CHASSIS CURRENT 02uAMEASUREMENT METHODDUT: OFF POL: REVGND:CLOSEFAULT CHASSIS CURRENT 15uAMEASUREMENT METHODDUT: OFF POL: FWDGND: OPENSAFETY ANALYZER SUMMARYPASS - WITHIN LIMITSSAFETY SPECS AAMIMAX GROUND RESISTANCE .15ohmMAX NORMAL GROUND CURRENT 500uAMAX FAULT GROUND CURRENT 1000uAMAX NORMAL CHASSIS CURRENT 100uAMAX FAULT CHASSIS CURRENT 300uAPASS / FAIL REPORT MESSAGESLNOTE:CThe Summary reports a Pass orFail message, telling which statecaused the failure.PASS - WITHIN LIMITSFAIL - PLUG POLARITYFAIL - LINE VOLTAGEFAIL - LOAD CURRENTFAIL - OPEN FUSEFAIL - NORM GND CURFAIL - FAULT GND CURFAIL - NORM CHS CURFAIL - FAULT CHS CUR