3-4GETTING READY FOR A TESTINFUSIONINFUSIONINFUSIONINFUSIONINFUSION ANANANANANALALALALALYZER--FLYZER--FLYZER--FLYZER--FLYZER--FLOOOOOWWWWWChoose Channel ScreenTo run a test, toggle to the appropriate Channel screenusing .STEP 2 9CHANNEL 1 IDLEPRIME WITH 50 mLTHEN PRESS STARTConnect and Prime DeviceConnect device under test (DUT) to the unit and prime thechannel. (See PRIMING for details.)STEP 1 9GETTING READY FOR A FLOW TESTNOTES:CThere can be up to two modulesconnected to the analyzer.CThe modules are interchangeable andthe unit will run with a single moduleconnected to either port.CThe Channel 1 or 2 screens will onlybe displayed when the modules areconnected to those ports.CThese instructions will utilize Channel1 for examples; Channel 2 functionsidentically.NOTES:CThe NORMAL LED on the remotemodule will blink at a rate of twice persecond, when a channel needs to beprimed.CAs the channel becomes ready, theNORMAL LED on the remote modulewill blink at a rate of once per second.CThe priming message will not displayif the DUT is already primed.