Document: 333D012G5-18-2007- 47 -Channel Stand Installation or Removal ProcedureTo install a stand, follow steps (1-7). To remove a stand, reverse steps (1-7)1) First gather all stand components and attaching hardware. The stand parts consist of two sides and oneback support. Hardware consist of (4) 1/4"x 1-1/4" bolts w/ 1/4" flat/lock washers and 1/4" hex nuts, (4)3/8" x 1 bolts w/ 3/8" flat/lock washers, (8) chrome dots, (4) Adjustable leg levelers. Tools required are a7/16” wrench or deep socket, 9/16” wrench or socket.2) Place stand sides on the floor so that the stainless steel legs are setting on the floor. Next place the standback support between the sides so that the open part of the channel is towards the floor. This can also bedone by rotating the stand parts (360 degrees), to make attaching the hardware easier.3) Using the 3/8 hardware, attach the sides and back support by inserting the bolts thru the (4) holeslocated in the support into the already installed rivnuts in the two sides. (2) bolts each side. Do nottighten any of these bolts until you have all (4) of them threaded in slightly. Tighten bolts after this isdone.4) With stand assembly complete, rotate stand so that the back of the sides are setting flat on the floor.Attach the leg levelers by screwing them into the bottom of the stainless steel legs in the sides of standassembly. Position the stand assembly in the upright position so that the leg levelers are now resting onthe floor. The stainless steel legs can be adjusted to your desired height requirement. This is done byremoving and replacing the (2) 3/8"bolts in each leg. Unused rivnuts in the legs will be plugged with the(8) chrome dots. If the cabinet is not in its permanent location you may want to perform the legadjustment after step 7.5) Lift the cabinet high enough to allow clearance for the stand assembly to be positioned under it. Liftingis typically done from the front of the cabinet. The stand assembly is slid in from the back of the cabinet.If space constraints are an issue, the stand may have to be assembled under the cabinet after it has beenlifted.6) Once the stand is positioned under the cabinet, install (4) 1/4"x 1-1/4" bolts into the key way slotslocated under the cabinet in each of the (4) corners. The cabinet is now ready to be lowered onto thestand. There are (4) holes on the top of the stand which the 1/4"x1-1/4" bolts must be guided thru. Becareful not to let the bolts drag thru the holes while lowering as burrs may occur and cause crossthreading. With the cabinet resting on top of the stand assembly, place (1) 1/4" flat / lock washer / hexnut onto each of the 1/4x1-1/4" bolts. Tighten after all (4) are threaded on.7) Lift device should be removed if not being used to move cabinet into position. If lift device is mobile itmay be used to maneuver cabinet into desired location. A blanket could also be placed under the leglevelers while sliding the unit to avoid scuffing the flooring.