8Ultraviolet Germicidal Lamp (optional)SterilGARDIII Advance can be equipped with an ultraviolet (UV) light. The light is controlled by anon/off switch on the front panel (see Figure 2, Operator Controls).WARNING• Do not defeat interlock, UV light is hazardous!• Eyes and skin should not be exposed to direct ultraviolet light.• Ultraviolet light should not be relied upon as the sole decontaminating agent. Additional surfacedisinfection should be performed both before and after every cabinet use.• A biological safety cabinet acts as a supplement to good aseptic practices, not as a replacement.UV lamps lose their effectiveness over time and should be replaced when the intensity at the worksurface drops below 40 microwatts per cm2 at a wavelength of 253.5 x 10-9 meters. Be sure to checkregularly.SpecificationsWeightThe weight of the SterilGARDIII, Model SG303 cabinet is 397 pounds with a shipping weight of 534pounds. With the optional stand, the cabinet weight is 457 pounds with a shipping weight of 594 pounds.Electrical Specifications115 VAC, 20A, 1 Phase, 60 HzSterilGARDIII Advance incorporates Baker’s StediVOLT motor speed controller for the blowermotor. This compensates for minor variations in incoming line voltages.SterilGARDIII Advance features one duplex GFCI protected receptacle protected by an internal 5Aself resetting circuit breaker. The unit is furnished with one 16' power cord with terminated with aNEMA 5-20 plug. The power cord provides the cabinet and internal receptacles with a protective earthground and serves as the main disconnect for the cabinet.The unit is protected with a 20 amp, 250 VAC time delay fuse.Blower motor (1,625 RPM) 3.20 ampsFluorescent Light (ballast) 0.49 ampsDuplex outlets (G.F.C.I.) 5.0 ampsTotal running load 8.69 amps