232. Press ENTER.3. Press ENTER when the unit displays “PrinterAttached?”4. When prompted with “Date Format” press theUP or DOWN arrow to show the date formatdesired. Press ENTER to select the desired for-mat and continue.5. When prompted with “Set Date,” press the UPor DOWN arrow to choose “Yes,” then pressENTER.6. Press the UP or DOWN arrow until you reachthe desired month. Press ENTER.7. Press the UP or DOWN arrow until you reachthe desired day. Press ENTER.8. Press the UP or DOWN arrow until you reachthe desired year. Press ENTER.9. When prompted with “Set Time” press the UP orDOWN arrow to choose “Yes,” then pressENTER. This is to set the time of day.10. Press the UP or DOWN arrow until you reachthe desired hour value. Press ENTER.11. Press the UP or DOWN arrow until you reachthe desired minute value. Press ENTER.12. Press BACK twice to return to the (Idle) mode.Setting the Cell ConstantThe actual temperature and cell constants are attached tothe cell cable internal to the NANOpure DIamond unit.This data helps ensure that the purity displayed is asaccurate as possible. The only time the user would needto enter this data is if the cell or main board is beingreplaced and/or updated.1. Shut power OFF to the unit and disconnect thepower cord.USER SETTINGS