267. Press ENTER.8. Press BACK to return to the (Idle) mode.The second line of the display will show a “Below Setpoint” message once every 8 to 10 seconds when themeasurement is below the set point. (The set point indica-tion is set at the factory at 10 MΩ-cm.)Resetting the Sanitization Timer1. From the (Idle) mode, when the display reads,“NANOpure DIamond (Idle) xx:xx:xxxx/xx/xx,” press the UP or DOWN arrow until thedisplay reads, “Reset Timers?”2. Press ENTER. Display will read, “Reset UVTimer (NO).”3. Press ENTER.4. Display will read, “Reset San Timer” and showon the right side of the second line when theuser will start receiving “Sanitize Unit”reminders.5. Press the UP or DOWN arrow to select “Yes.”6. Press ENTER and the timer will be reset. Thiswill reset the sanitization timer for approximatelysix months calendar time.Resetting the UV TimerWhen the NANOpure DIamond Life Science (UV/UF) is inan (Idle) position:1. From the (Idle) mode, when display reads,“NANOpure DIamond (Idle) xx:xx:xxxx/xx/xx,” press the UP or DOWN arrow until thedisplay reads, “Reset Timers?” Press ENTER.2. Press the UP or DOWN arrow until the displayreads, “Reset UV Timer.”USER SETTINGS