3. To accept change, press Confirm softkey.• Setting changes and System Config- Syringe page 2 displays.2.4.7 Near EndEnabling Near-End-of-Infusion ( NEOI)allows an alert to be set from 1 to60 minutes before the infusion is complete.The alert occurs at the configured time orwhen 25% of the VTBI remains, whicheveroccurs later.NOTE: In the following procedure, the NearEnd option is changed to Enable. Changingthe option setting to Disable disables theChange Time soft key, as displayed instep 1 .1. After accessing System Config -Syringe options display, page 2 , pressNear End soft key.2. To change option setting, press soft keynext to applicable option (Enable orDisable).3. To change Alert Time, press ChangeTime soft key and enter new time(1 to 60 minutes).4. To accept changes(s), press CONFIRMsoft key.• Settings change and System Config- Syringe page 2 displays.2.4 CONFIGURATION SETUP - SYRINGEMODULE (Continued)2.4.6 Max rate (Continued)CHECKOUT AND CONFIGURATION2-7Alaris ® Syringe Module (8110 Series)Alaris ® PCA Module (8120 Series)Technical Service ManualDisableDisableCONFIRMSystem Configuration - SyringeChangeTimeDisableEnableNear-End-ofInfusionAlert Time:2 minAllows use of the Near-End-ofInfusion feature.DisableDisableCONFIRMSystem Configuration - SyringeChangeTimeDisableNear-End-ofInfusionAlert Time:2 minAllows use of the Near-End-ofInfusion feature.EnableDisableDisableCONFIRMSystem Configuration - SyringeDisableNear-End-ofInfusionAlert Time:10 minAllows use of the Near-End-ofInfusion feature.EnableChangeTimeSystem Configuration - SyringeMaximum rateadjustmentMaximumrate:mL/h99.9ChangeValueConfirm