Datacolor CONDITIONER™ (MB2) 06/2017 page 72/12610.6.4 Setpoint entry• Select the field “Temperature” and enter the desired temperature setpoint.Cabinet without lights: Setting range: -5 °C up to 70 °C.Cabinet with lights: Setting range: -5 °C up to 50 °C.Confirm entry with Confirm icon. The controller returns to the section view.• Select the field “Humidity” and enter the desired humidity setpoint.Cabinet without lights: Setting range: 0% r.H. up to 80% r.H.Cabinet with lights: Setting range: 0% r.H. up to 70% r.H.Confirm entry with Confirm icon. The controller returns to the section view.• Select the field “Fan” and enter the desired fan speed setpoint.Setting range: 40% up to 100% fan speed.Confirm entry with Confirm icon. The controller returns to the section view.• Cabinet with lights: Select the field “Light” and enter the desired light intensity setpoint.Setting range: 0% up to 100% light intensity. With setting 0% the light is off.Confirm entry with Confirm icon. The controller returns to the section view.10.6.5 Special controller functions via operation linesYou can define the switching state of up to 16 operation lines (control contacts). They are used to activate/ deactivate special controller functions.• Operation line “Humidity off” serves to turn off the humidity.• Operation line Idle mode” activates / deactivates the operating mode “Idle mode”.The other operation lines are without function.Select the desired program and section. You can set the operation lines in the “Functions on/off” field.For details please refer to chap. 9.7.3.