Datex-Ohmeda S/5 Light Monitor1.5 Symbols1.5.1 Symbols on equipmentAttention, consult accompanying documents.on the rear panelindicates thewarning:Electric shock hazard. Do not open monitor frame. Refer servicing to qualified personnel.Do not touch any part of monitor or patient connections during defibrillation procedure.Disconnect power supply before servicing. Use only external power adapter N-LPOW.on the rear panel dustfilter indicates thereminder:Check rear panel dust filter regularly, and clean at least once a month.on the patientconnector panelindicates thecautions:Use only the specified cables and accessories.Ensure proper contact of the return electrode of the electrosurgery unit to your patient toavoid possible burns at ECG electrode or other probe sites.on the external poweradapter indicates thecautions:For continued protection against fire hazard, replace only with same type and rating of fuse.Make sure voltage selector is set according to local voltage.For indoor use only. For patient transport outdoors use the monitor in protective case andbattery powered.Do not immerse in liquid or allow any liquid to enter the interior.on the battery unitindicates thewarnings andcautions:Do not put in fire.Do not short-circuit any terminals.Do not disassemble or mutilate.Use only with Battery Module or External Battery Charger.Discharge and charge batteries fully every 3 months or more often, if required, to restore fullcapacity.Sealed NiCd battery must be recycled or disposed of properly. Discharge before disposing.on the CPU boardindicates the caution:There is a lithium battery on the CPU board. Discard useless battery according to localregulations.on the CO2 sensor Do not immerse the sensor in liquids or autoclave it.Document No. M10513394