Datex-Ohmeda S/5 Light MonitorDocument No. M105133982 SAFETY2.1 Safety precautions2.1.1 WarningsWARNING Indicates situations in which the user or the patient may be in danger of injuryor death.Explosion hazard• Do not use the monitor in presence of flammable anesthetics.Electrical shock hazard• Use only hospital grade grounded electrical outlets and power cord. Grounding reliabilitycan only be achieved when the equipment is connected to a receptacle marked ‘hospitalgrade’.• Do not remove cover. Refer servicing to qualified service personnel.• Avoid any conductive contact to the probes and sensors.• Disconnect the monitor from the electrical outlet before cleaning. Let it dry completelybefore reconnecting it to the electrical outlet.• Do not autoclave any part of the monitor with steam or ethylene oxide. Do not immerse inliquid or allow liquid to enter the interior of any part.• Make sure that external equipment is hospital grade grounded before connecting it to themonitor. Do not connect any external equipment to the monitor, except that specified by GEHealthcare.• When connecting external equipment to the monitor, make sure that the whole combinationcomplies with safety standard IEC 60601-1-1 for the medical electrical systems and withthe requirements of local authorities. Non-medical electrical equipment connected to themonitor shall be supplied from an extra isolating transformer which has a protectivelyearthed outlet.• Do not leave loose any sensor cables connected to patient. Either connect them to themonitor or detach them from the patient.• Do not use the monitor, any parts thereof or any accessories, if they have become wet orexposed to liquid.• Where the integrity of the external protective conductor installation or arrangement is indoubt, equipment shall be operated from its internal power source.