Service check6.2.2 Recorder test10. Open the paper compartment cover. Check that the message “Recorder: Cover open”appears on the screen, then close the cover.11. Select Main Menu - Record/Print- Record Waveand check that the recorder starts recording. Select Stop Wave to stop recording.Check that the quality of the recordings is acceptable."6.2.3 PCMCIA card test12. Insert the Memory card labeled “Data” in the DataCard slot. Check that the message “DataCard inserted” appears in the message field and the white data card symbol in the upperright hand corner of the screen appears within one minute."6.2.4 Network test13. Connect the Mon-Net cable and the Identification plug to the monitor.14. Check that the monitor connects to the Datex-Ohmeda Network, i.e. the network symbolis displayed on the upper right-hand corner of the screen.A message regarding the connection to Datex-Ohmeda Information Center should appear inthe message field on the screen.15. Press the Menu button to enter the Network service menu:Monitor Setup - Install/Service (Password 16-4-34) -Service View (Password 26-23-8) - Monitor - CommunicationCheck that:− the “location ID number” matches with the ID plug connected to X3− the packets and bytes IN is increasing slowly− the packets and bytes OUT is increasing quickly.− the “connections” shows the names of the connected networks− the counters for data errors (“CRC”, “Frame”, “Transm.”) are stableNOTE: The counters may show values greater than 0. However, any values increasingcontinuously indicate a problem.16. Check that the counters for hardware errors (“Intern.”, “Missed”, “FIFO”, “Overrun”) all show0. If any of the counters show a value greater than 0, it indicates a problem on the CPUboard.Return to the Monitor Setup menu by selecting Previous Menu repeatedly."Document No. M105133935