Cleaning and MaintenanceRepairsRepair work must be performed by qualified factory-trainedpersonnel only.Warning!Never lift balance by the weighing pan as this may causedamage to internal mechanisms.Always lift and transport the balance by its base.CleaningCaution! Disconnect the balance AC adapter from power source prior tocleaning. Make sure that no liquids enter the balance housing. Do not useaggressive cleaning agents such as cleansers. A mild detergent is recom-mended. Disassemble the Pan Assembly and clean the floor pan, breezering, pan support and cover pan separately, then reassemble. Clean the bal-ance with a piece of cloth. After cleaning, wipe the balance down with a dry,soft cloth. Recalibration of the balance is recommended after cleaning.Warning!If there are any indications that safe operation of the balance isno longer warranted, turn off power and disconnectfrom AC power source immediately.Safety InspectionSafe operation of the balance is no longer assured if there is visible damageto the AC adapter or cord, the AC adapter no longer functions properly orthe AC adapter has been stored for a long period under unfavorable condi-tions.NoteThis unit contains no user serviceable parts.All replacement parts should be obtained from themanufacturer. Please refer to the inside front cover ofthis manual for the phone number of your sales andservice representative.!!36