46SpecificationsVerified Models with EC Type ApprovalModel TB-215DType BC BC 100Accuracy class kMaximum capacity, Max* g 60/210Scale interval, d* mg 0.01/0.1Verification scale interval, e* g 0.001Minimum capacity, Min* g 0.001Tare range (subtractive) g –210Range of use according to Directive No. 90/384/EEC* g 0.001–210Response time (average) s ≤12/3Adaptation to ambient conditions and application By selection of 1 of 4requirements optimized filtersDisplay update (depends on the filter level selected) s 0.2–0.4Allowable ambient operating temperature °C +5…+40Operating temperature range °C +15…+25Pan size mm Ø 80Weighing chamber height (effect. dimens.) mm 225Net weight approx. kg 6.6Selectable weight units g, mgSelectable application programs Mass unit conversion by tog-gling; tare memory; net total;weighing in percent; animalweighingAutomatic zero tracking function Standard feature(can be turned off by menu code)Power requirements V~ Via AC adapter, 230 VAC or115 VAC, –20%…+15%Frequency Hz 48–63Power consumption (average) VA 16: maximum; 8: averageHours of operation with fully charged optional h 25external battery pack, approx.:Built-in interface port RS-232 C-S/V24-V28; 7-bit;parity: even, mark, odd, space;transmission rates: 150…19,200 baud; 1 or 2 stop bits;software/hardware handshakeStandard features/equipment supplied:Dust cover xAC adapter, varies acc. to country xBuilt-in calibration weight xLevel indicator x* = EC Directive No. 90/384/EEC on non-automatic weighing instruments applicable inthe E.U. and within the European Economic Area