4INM MTL 130-0202 Rev 42.11 Analogue outputs4 to 20mA – programmable as follows:Range Gas measured Low End High EndOption 1 Range 1 H 2 4mA equiv. to 0% 20mA equiv. to 1.00%…5.00%- user adjustableRange 2 Cl 2 4mA equiv. to 0% 20mA equiv. to 100% - fixedOption 2 Range 1 H 2 4mA equiv. to 0% 20mA equiv. to 2.00%…10.00%- user adjustableRange 2 Cl 2 4mA equiv. to 0% 20mA equiv. to 100% - fixedMaximum load resistance on outputs: 600 ohms2.12 Fault indicator outputsTwo fault-indicating relay outputs are provided:1. Instrument status2. Reactor status ±10°C of set point.Both are volt-free SPDT relay contacts rated at 30V AC or DC, 0.5A. The “no-fault” conditionis NC - see Figure 6.2.13 Concentration alarmsTwo user-configurable Alarm outputs are provided:1. Programmable alarm for Cl 22. Programmable alarm for H 2Both are volt-free SPDT relay contacts rated at 30V AC or DC, 0.5A. The “no-alarm”condition is NC - see Figure 6.2.14 Speed of responseT90: 30 seconds typical2.15 Power requirementsVoltage: 99...132V or 198…264V AC, 50/60HzPower: 80VA2.16 Electronics unitMounting: Wall mounting caseDimensions: 650mm (H) x 450mm (W) x 250mm (D) - without glandsMaterials: Polyester enclosure -(IP66 rating when door is closed and with suitable cable glands)Net weight: 20 kg (nom.)2.17 Katharometer/reactor unitMounting: Wall mountingDimensions: 150mm (h) x 240mm (w) x 115mm (d) including heat shieldKatharometerCable:2 metre length provided -cable must not be extended or certification is affected.Net weight: 3 kg (nom.)