Page 565.2.4 System Fuse SpecificationsTable 6 below summarizes the location and specification of the fuses used in BioTector. The locations of thefuses are also displayed in figures 9 and 10 above.Table 6 System Fuse Specifications230 VSystems115 VSystemsLocationNamePCB IDNumberInterruptRatingType Material FuseNumberCurrentRatingCurrentRatingPower PCB(Mains PCB) 81204030-02 H-250VMiniature5x20mm CeramicF1 T 1.25A T 2.50AF2 T 500mA T 500mAF3 T 1.00A T 1.00AF4 T 1.60A T 2.50AF5 T 2.00A T 3.15AF6 T 500mA T 500mARelay PCB 81204001-01 L-250VMiniature5x20mm GlassF1 T 2.50A T 2.50AF2 T 500mA T 500mAF3 T 500mA T 500mAF4 T 1.00A T 1.00AF5 T 500mA T 500mAF6 T 1.00A T 1.00AF7 T 1.00A T 1.00AF8 T 1.00A T 1.00AStreamExpansionPCB81204040-02 L-250V Miniature5x20mm Glass F1 T 1.00A T 1.00ASignal PCB 81204010-02 L-250V Miniature5x20mm Glass F1 T 1.00A T 1.00AF3 T 500mA T 500mAMain Board(Motherboard) 81204022-02 L-250V Miniature5x20mm Glass F1 T 500mA T 500mACooler DINRail Terminal 47 L-250V Miniature5x20mm Glass F1 T 2.5A T 2.5A24V DCCirculationPump DINRailTerminal 42 L-250V Miniature5x20mm Glass F1 T 2.5A T 2.5ANP I/O PCB(TNTP Board) 81204290-01 L-250V Miniature5x20mm GlassF2 T 630mA T 630mAF4 T 1.00A T 1.00AF6 T 1.00A T 1.00AKEY A: Amperes DIN: German Institute for Standardization (Deutsches Institut für Normung e.V.)F: Fuse H: High Interrupt ID: Identification L: Low InterruptmA: Milli-amperes PCB:Printed Circuit Board T: Time Lag (Time Delay)V: VoltsBioTector contains electrical components operating under high voltages.Contact may result in electric shock and severe or fatal injury.All electrical work should be carried out by qualified electrical personnel only.When any fuse replacement is required in the system, please refer to table 6below.DANGER