14SpecificationsUser interfaceTouch-screen panel Displays concentration, trend graph, diagnostics,alarm status, historical data.Password protection Five levels of authorised access to configuration anddata managementSensorSample temperature Measurement from -5 to 50°C (23 to 122°F)Sensor resistant to temperature from -5 to 100°C (23 to 212°F)Sample pressure 1 to 20 bar absolute (14.5 to 290 psia)Sample flow rate requirement 50 to 300 mL/minSample types Water onlyCIP Not designed for use in CIP processesMeasurement range Low range sensors: 0 to 2,000 ppb (dissolved. Indicatives values until 5000 ppb)High range sensors: 0 to 40 ppm (dissolved)Repeatability Low range sensors: ± 0.4 ppb or 1%, whichever is the greaterHigh range sensors: ± 0.015 ppm or ± 2% whichever is the greaterReproducibility Low range sensors: ± 0.8 ppb or 2%, whichever is the greaterHigh range sensors: ± 0.02 ppm or ± 3% whichever is the greaterAccuracy Low range sensors: ± 0.8 ppb or 2%, whichever is the greaterHigh range sensors: ± 0.02 ppm or ± 3% whichever is the greaterLimit of detection (LOD) Low range sensors: 0.6 ppbHigh range sensors: 0.015 ppmResponse time (90%) Low range sensors: < 10 seconds in gas phase; < 30 seconds in liquid phaseHigh range sensors: < 10 seconds in gas phase; < 50 seconds in liquid phaseWarm-up time NoneDisplay resolution 0.1 ppbCalibration Low range sensors: Single point calibration (zero)High range sensors: Two at cap replacement (zero and air), one during use (air)Calibration sample Low range sensors: Standard 99.999% N2 (quality 50) or equivalent oxygen free gasHigh range sensors: Standard 99.999% N 2 (quality 30) or equivalent oxygen free gas, airAccessoriesAccessoriesActive spots, spare sensors, tool kit, portable calibration setup, etc.Ask your local Hach representative for more details on all available spare parts andaccessories.InstallationIn process sensor installationdevicesORBISPHERE insertion/extraction valve for installation on Varinline access unitsORBISPHERE 28 mm sensor weld-on socket28 mm stationary housing for installation on Varinline access unitsAmbient temperature -5 to 50 °C (23 to 122 °F)Humidity 0 to 95% non-condensing relative humidityPower supply Universal 85-264 VAC @ 50/60 Hz, 25 VA10-36 VDC, 25 WWeightsWall and pipe mount(H x D x W)236.5 x 160 x 250 mm - weight 4.25 kg9.31 x 6.30 x 9.84 in. - weight 8.82 lbsPanel mount: Face (housing)(H x D x W)156 (123) x 250 x 220 (214) mm - weight 3.35 kg6.14 (4.84) x 9.84 x 8.86 (8.43) in. - weight 6.62 lbsK1200 28 mm sensor 0.74 kg (1.63 lbs)Calibration device 0.7 kg (1.54 lbs)