Figure 8 High-voltage barrierWiring for powerD A N G E RElectrocution hazard. Protective Earth Ground (PE) connection is required.D A N G E RElectrical shock and fire hazards. Make sure to identify the local disconnect clearly for the conduitinstallation.W A R N I N GPotential Electrocution Hazard. If this equipment is used outdoors or in potentially wet locations, aGround Fault Interrupt device must be used for connecting the equipment to its mains power source.W A R N I N GElectrocution hazard. The local disconnection means must disconnect all the electrical current-carryingconductors. Mains connection must keep supply polarity. The separable plug is the disconnect meansfor cord connected equipment.W A R N I N GElectrical shock and fire hazards. Make sure that the user-supplied power cord and non‐locking plugmeet the applicable country code requirements.W A R N I N GExplosion hazard. This manual is only for installation of the unit in a non-hazardous location. Forinstallation of the unit in hazardous locations, use only the instructions and approved control drawingprovided in the hazardous location installation manual.N O T I C EInstall the device in a location and position that gives easy access to the disconnect device and its operation.English 15