Orbisphere 510/511/512 Analyzers - Installation 17 of 138Operator Manual ORBISPHERE1.4.2 MountingFig 1-8: Panel Mount Bracket Frame1) Cut an opening in the panel toaccommodate the bracket frameprovided (this is the same size asprevious generations of Orbisphere type3600 instruments).Fig 1-9: Opening Dimensions2) Install the provided frame in the opening3) Fold the 6 tabs over the panel lips, usingadjustable joint pliers.4) Slide the instrument in the bracket frame.The instrument should go over the four“T” pins. Rotate the 4 fast locking screwson both sides of the front panel and slideit in.5) Rotate the 4 fast locking screws 1/4 turntwice in the lock direction as indicated onthe side of the front panel. This locks theinstrument in place on the four “T” pins.6) To access the connections inside theinstrument, remove the instrumenthousing (six screws on the back panel,and slide the housing back out)7) Pass the cables through the housing,then through the cable gland (ifapplicable) and then perform theconnections as detailed below.Note:Do not forget to pass the cable through thehousing before passing the cable through thecable gland on the back panel.