88 of 138 Communication Menu - Orbisphere 510/511/512 AnalyzersOperator ManualORBISPHERE7.2 PROFIBUS-DP Communication (optional)7.2.1 InstallationOn the Orbisphere CD, there is an “Orbi2079.gsd” and an “Orbi2079.bmp” file availablein the “Profibus DP” folder to help configure the PROFIBUS-DP. The GSD file containsthe following elements:• A module to decode the barometric pressure value and unit• A module to convert the channel measurement data such as gas concentration,gas unit, temperature, temperature unit and the events.WARNINGInstallation should be performed exclusively by personnel specialized and authorized towork on electrical installations, in accordance with relevant local regulations. Disconnectthe power supply of the instrument before carrying out any work inside the instrument.CAUTION:Proper ESD (electrostatic discharge) protocols must be followed to prevent damage tothe product. All fittings must be properly seated and tightened to prevent any water anddust ingress.1) Install the PROFIBUS-DP module and the jumper J3 on the main board (locationhighlighted in the illustration below).