Handicare Freecurve74The LED remains dim upon joystick operation in the faulty directionConnect the short-circuit connectors K6 and K9A. The lift moves after operation of the direction switch (DS)The fault is located in the chair• Wire break between K6-3 and the joystick• Wire break between K6-5 and the joystick• Joystick is defect• Earth short at K6-3, K6-5B. The lift does not move after operation of the direction switch (DS)The fault is located in the unit• Wire break between K6-3 and K4-11• Wire break between K6-5 and K4-10• Directions switch DS is faulty• Earth short at K4-10, K4-11The LED flashes upon joystick operation in one directionConnect the short-circuit connectors K6 and K9A. The lift moves after operation of the direction switch (DS)The fault is located in the chair• One of the footrest edges is activated• One of the footrest switches is faulty• The seat frame bottom detection is activated• The seat frame bottom switch is faulty• Wire break between K9-1 and K9-2• Wire break between K9-3 and K9-4B. The lift does not move after operation of the direction switch (DS)The fault is located in the unit• One of the cover safety edges is activated• The cover detection switch is faulty• Wire break between K4-4 and K9-3• Wire break between K4-5 and K9-4• Wire break between K4-6 and K9-1• Wire break between K4-7 and K9-29 | The lift does not move in one direction10 | The lift does not move in one direction