13CommissioningCheck whether the voltage given on the type plate cor-responds to the available mains voltage�The socket used must be earthed(fitted with earth contact)�If these conditions have been met, the machine is readyfor operation when the mains plug is plugged in�If these conditions are not met, safe operation is not guar-anteed and the machine could be damaged�Observe the ambient conditions (temperature, humidity,etc�) listed under “Technical Data”�The battery in the Wireless Controller (WiCo) must becharged at the station before the first use of the device�Mounting the Wireless Controller to the station:Put the WiCo into the holder at the station while pressingthe release button�After several seconds, screen display the information of WiCo�IKA®CirculatorIC/HBC ControlVersion 1�3�000Then the working screen in standby status appears andthe device is ready for operation�InformationInternal Max: 95 °CInternal Min: 5 °CExternal Max: 95 °CExternal Min: 5 °CFluid H2 O (Water)A0�00int °Cext °Cint 0�00 1000set °C Pump rpmSafe Temp °CChange the temperature setting with left knob (K)�Change the pump speed setting with the right knob (J)�In standby status, press the left knob (K)� The device startsheating function and the pump begins to run�In work status, press the right knob (J) to stop operationof the pump� The heating function and the pump stops�Note: In standby status, press the right knob (J) to startthe pump function, the heating function will not be acti-vated�In working status, press the left knob (K) to stop the heat-ing function, the pump keep running�When the WiCo is switched on without being connectedto the station, the green LED bar (D) and the green Blue-tooth® LED (F) on circulator lights up�The controls of the WiCo allow locked by pressing the keybutton (O), so no accidental changes during operation arepossible (key symbol appears in the display) �By pressing the key button (O) again, the controls are re-leased (key symbol disappears from the display)�Fig. 18Fig. 19Fig. 20If the WiCo should be permanently attached to the station,we strictly recommend to fasten the unlocking button withthe integrated screw (turn counter clock wise)�After switch on the mains switch (2) at the rear side ofthe station, the screen of the WiCo displays the devicedesignation and the software version after a beep�Fig. 17