Part No. 960-000165R_Rev. 1 © 2013, JCM American, Corporation1 8Lecture NotesVEGA™ Banknote Acceptor JCM Training Overview July, 2013NON-PC CALIBRATION PROCEDURESWhen removing or replacing the Sensor, Sub-Sensor, or CPU Board, performthe following steps to calibrate the Entrance, Validation, and Edge Sensors ofthe VEGA Unit.1. Figure 11 illustrates the Calibra-tion connection Tools and Har-ness configurations necessarywhen calibrating WITHOUTusing a PC.2. Set DS 1, Switches #1, #2, & #8to ON. Ensure ALL DS 2Switches are OFF.3. Apply power to the VEGA andthe Front Panel LED will blinkGreen.4. Set DS 1, Switch #1 OFF.5. The Front Panel LED will nowslowly blink at a Cyan Colorrate.6. Set DS 2, Switch #1 ON.This action begins the Entrance Sensor's Adjustment Procedure withoutReference Paper in Place.7. The Front Panel LED will slowly blink at a White Color Rate once the EntranceSensor's adjustment is complete.NOTE: If any of the Sensors are dirty the calibration will not beaccurate. Be sure to clean all the Sensors before calibratingthe Unit.NOTE: There are two types of Calibration Procedures:• Calibration using a PC, and• Calibration without using a PC.To calibrate using a PC, refer to the VEGA Operation andMaintenance Manual, Section 6, Page 6-6.Figure 11 Non-PC Calibration Toolsand CablingaedbInterface ConnectorPower SupplyTo ACReceptaclePowerca) VEGA Unitb) Power Supply (e.g., UAC, VM-VEGA, etc.)c) Harnessd) White Reference Paper KS-325 (Part#190876)e) Black Reference Paper KS-326 (Part #190875).NOTE: Do not touch theVEGA Unit while it's adjusting.NOTE: Keep DS 2 Switches ON until instructed to turn them OFF.