11 Instant helpIn case of an error in the program process, briefly turn off the balance and disconnectfrom power supply. The weighing process must then be restarted from the beginning.Help:Fault Possible causeThe displayed weight does not glow. • The balance is not switched on.• The mains supply connection has beeninterrupted (mains cable not pluggedin/faulty).• Power supply interrupted.• Batteries are inserted incorrectly or empty• No batteries inserted.The displayed weight is permanentlychanging• Draught/air movement- • Table/floor vibrations• The weighing plate is in contact withforeign matter.• Electromagnetic fields / static charging(choose different location/switch offinterfering device if possible)The weighing value is obviously • The display of the balance is not at zerowrong • Adjustment is no longer correct.• Great fluctuations in temperature.• Electromagnetic fields / static charging(choose different location/switch offinterfering device if possible)Should other error messages occur, switch balance off and then on again. If the errormessage remains inform manufacturer.440-BA-e-094132