ABP-BA-e-1810 10515.5 Data transmission format1. Standard format example [-123.4567 ] Positon 1 2 3 4 5 6 7 8 9 10 11 12 13ASCII 2DH 31H 32H 33H 2EH 34H 35H 36H 37H 20H 67H 20H 0DHData - 1 2 3 . 4 5 6 7 g C/RNo. Description Signs [ ˽ ] positive values (blanks)[ - ] negative values Weighing Value Numeric weighing value is displayed in eight positions.Not required positions = space character 20HPossible overload, presented in 2 positionswith O L.For scales with type approval the verified value ispresented in brackets "[ ]". This way the data length isincreased by two positions. Unit 1 character: Position 123 character: Position 11-134 character: Position 11-14 Final character Separator C/R = 0DH, L/F = 0AHAt CR+LF data length will be increased by one position.2. Stable / unstable weighing value:Positon 1 2 3 4ASCII 53H 2DH 31H 32HData S - 1 2stable S (53H)unstableU (55H)