110 ABP-BA-e-181015.7 Communication parametersAll communication parameters will be set (See chap. 15.7.1) by calling a standardsetting.The subsequent standard setting must be selected according to the printer (detailssee the following table).All parameters may be of course also set in a user specific way (See chap. 15.7.2).Menuselection Standard Extended Type M Type S Type A User settingManufacturerShimadzu(Standard)Shimadzu* Mettler Sartorius A-D -- Settingfor KERNYKB-01NBaudRate 1200 1200 2400 1200 2400 user-defined 9600Parity None (8) None (8) Even (7) Odd (7) Even (7) user-defined None (8)Stop bit 1 1 2 2 2 user-defined 1Hand-shake Hardware Hardware off Hardware off user-defined offDataformatShimadzuStandardShimadzuStandardMettlerStandardSartoriusStandardA-DStandarduser-defined FREESeparator C/R C/R C/R + L/F C/R + L/F C/R + L/F user-defined C/R*only if the balance can send a feedback to the PC (without error: OK [C/R], at error NG [C/R].15.7.1 Select standard setting1. Call functionPress and hold PRINT-key for approx. 3 sec.Use the navigation keys to select and confirm usingthe OK-key.Use the navigation keys to select interfaceand confirm using the OK-key.