28 ABP-BA-e-18108.2 Time-controlled automatic adjustmentWith the help of its internal adjusting weight and integrated clock the balance can beset to carry out automatic adjustment at set times (up to three times daily). Thisfunction is a very convenient function, when adjustment reports are desired to bemade for regular adjustments, or when wishing span adjustments during break timesto avoid interruption of measurement work.The weight symbol blinks for about two minutes as notification of span calibrationbefore it begins. Automatic adjustment can be stopped by actuating the [ON/OFF]key during this message.Parameter setting:Press and hold the CAL-key for approx.3 sec. and the menu for setting will be displayed.Use the navigation keys to select CAL> and confirm using the OK-key.Select first time and confirmusing the OK-key.Select setting [on] or [off] and confirmusing the OK-key.Select [Setting] to enter time.Use the navigation keys to enter timeand confirm using the OK-key.Repeat these operating steps to set thetime for / .