4.2 Entering Density for FluidPress the [CAL] key repeatedly until "SettinG" appears.Press the [TARE] key.Repeatedly actuate the [CAL] key until "LSG SEt" appears(Example)Press the [TARE] key. The currently set density for the liquid to bemeasured appears. In the upper part of the display panel, thesymbol and the # symbol appear in order to indicatenumerical input status. The leftmost digit blinks.(Example)Enter density for the liquid to be measured. When the [UNIT] keyis pressed, the numerical of the blinking digit increases by 1 at atime. You can determine the value of the flashing digit, or shift theflashing digit by one position to the right, by pressing the [PRINT]key. Confirm your setting by pressing the [TARE] key.Repeatedly press the [ON/OFF] key until the balance is inweighing mode.ABT-A01-BA-defsi-0710 33