4.3 Carrying Out MeasurementRepeatedly press the [UNIT] key until the balance is in densitydetermination mode for solids ",d". Note that “g” also appearsduring weight measurement in air.Press the [TARE] key. Place the object to be measured on thesample dish.When standstill control is complete, press the [CAL] key.Place the item to be measured on the immersed sifting bowl. Thedisplay is showing the density of the measured item. “dSP oL” maybe displayed when nothing is on the weighing pan, which isnormal.To start next measurement, press the [CAL] and the [ TARE ] keyand place the object to be measured on the weighing tray.5 Determining density of liquidsPrepare balance as described in chapter 2 "Installation of density determination set".Ö Install holder for the thermometer on beaker rim.Ö Suspend thermometerÖ Fill your measuring liquid into the beaker. Filling height should be approx. ¾ of thecapacity.Ö Heat measuring liquid until temperature is constant.Ö Prepare glass sinker5.1 Activate functionPress several times the [UNIT] key to change the display between activated units,piece counting, percentage and density determination mode.This does not require additional software.34 ABT-A01-BA-defsi-0710