10 Instant helpIn case of an error in the program process, briefly turn off the balance anddisconnect from power supply. The weighing process must then be restarted fromthe beginning.Help:Fault Possible causeThe displayed weight does not glow. • The balance is not switched on.• The mains supply connection has beeninterrupted (mains cable not pluggedin/faulty).• Power supply interrupted.• Batteries are inserted incorrectly or empty• No batteries inserted.The displayed weight is permanentlychanging-• Draught/air movement• Table/floor vibrations• The weighing plate is in contact withforeign matter.• Electromagnetic fields / static charging(choose different location/switch offinterfering device if possible)The weighing value is obviouslywrong• The display of the balance is not at zero• Adjustment is no longer correct.• Great fluctuations in temperature.• Electromagnetic fields / static charging(choose different location/switch offinterfering device if possible)Should other error messages occur, switch balance off and then on again. If the errormessage remains inform manufacturer.30 EMS-BA-e-1514