34 EOC-BA-e-17117.9 Manual totalizingWith this function the individual weighing values are added into the summationmemory by pressing and edited, when an optional printer is connected.• Menu setting:„P2 COM“ „MODE“ „PR2“, see chap. 8.2• The totalizing function is not active when the weight is under 20d.Add up:Place load A.Wait until the stability display appears, then press. „ACC 1“ will be displayed shortly, then theweighing value will be displayed. The weight value will besaved and printed if an optional printer is connected. Thesymbol for Total will be displayed.(example)Remove the load. More loads can only be added whenthe display ≤ zero.Place load B.Wait until the stability display appears, then press .„ACC 2“ is shortly displayed. The weight value will beadded into the summation memory and printed ifnecessary.(example)Add more loads as described before.Please note that the weighing system must be unloadedbetween the individual weighing procedures.This process may be repeated 99 times or till such timeas the capacity of the weighing system has beenexhausted.