Wait for stability display „STABLE“, then press . When“LoAd 3“ is displayed, put the third adjustment weight (1/3max) carefully in the centre of the weighing platform. Wait for stability display „STABLE“, then press . After the adjustment the balance will carry out a self-test.Remove adjusting weight during selftest, the appliance willreturn into weighing mode automatically.An adjusting error or incorrect adjusting weight will beindicated by the error message; repeat adjustmentprocedure.6.9 VerificationGeneral introduction:According to EU directive 90/384/EEC or 2009/23EG balances must be officiallyverified if they are used as follows (legally controlled area):a) For commercial transactions if the price of goods is determined by weighing.b) For the production of medicines in pharmacies as well as for analyses in themedical and pharmaceutical laboratory.c) For official purposesd) For manufacturing final packagesIn cases of doubt, please contact your local trade in standard.Verification notes:An EU type approval exists for balances described in their technical data asverifyable. If a balance is used where obligation to verify exists as described above, itmust verified and re-verified in regular intervals.Re-verification of a balance is carried out according to the respective nationalregulations. The validity for verification of balances in Germany is e.g. 2 years.The legal regulation of the country where the balance is used must be observed!Verification of the balance is invalid without the seal.The seal marks attached on verified balances point out that the balance mayonly be opened and serviced by trained and authorised specialist staff. If theseal mark is destroyed, verification looses its validity. Please observe allnational laws and legal regulations. In Germany a re-verification will benecessary.16 KFE-TM-BA_IA-e-1313