63 PBS/PBJ-BA-e-131416.5 Auto Print function(models with non-verifiable setting only)Using the Auto Print function ensures automatic data printing without necessity topress the PRINTbutton for each single measurement. If the function is activated, the(Auto-Print) symbol is displayed.It is possible to select from six types of automatic printing. Information concerningsetting the zero range can be found in chapter 11.7Printing whilst loading:Select the menu element 42 in the menu overview.Place or remove the sample when the displayed value lies within the zero range. Thedata will be output automatically when the stabilisation symbol is illuminated andthe displayed positive value exceeds 5 times the zero range value. The next dataoutput will occur when the display is brought to the value within the zero range byremoving the sample or pressing the TARE /button.Printing whilst loading and unloading:Select the menu element 43 in the menu overview.Place or remove the sample when the displayed value lies within the zero range. Thedata will be output automatically when the stabilisation symbol is illuminated andthe displayed positive or negative value exceeds 5 times the zero range value. Thenext data output will occur when the display is brought to the value within the zerorange by removing the sample or pressing the TARE /button.Printing whilst loading and at zero:Select the menu element 44 in the menu overview.Place or remove the sample when the displayed value lies within the zero range. Thedata will be output automatically when the stabilisation symbol is illuminated andthe displayed positive value exceeds 5 times the zero range value. Remove thesample or press the TARE /button. The data will be output again when the displayedvalue lies within the zero range and the stabilisation symbol is illuminated.Printing whilst loading, unloading and at zero:Select the menu element 45 in the menu overview.Place or remove the sample when the displayed value lies within the zero range. Thedata will be output automatically when the stabilisation symbol is illuminated andthe displayed positive or negative value exceeds 5 times the zero range value.Remove the sample or press the TARE /button. The data will be output again whenthe displayed value lies within the zero range and the stabilisation symbol isilluminated.