PBS/PBJ-BA-e-1314 7617.4.3 FormatSpecify the format of data output by the balance.For the standard format of the KERN electronic balances• Select menu element 77 .For the old data output format of the KERN electronic balances:• Select menu element 78 .17.4.4 Speed of communicationSpecify the transmission speed (300, 600, 1200, 2400, 4800, 9600, 19200 or38400 bps).The value presented as „b-xxx” gives the number of bps (bits/second). Baud-rateand bps are the same value.Select one of the menu elements from 81 to 88.17.4.5 Parity / Bit lengthSelect the parity and bit length.No parity, 8-bit length: Select the menu element 89Odd parity, 7-bit length: Select the menu element 90Straight parity, 7-bit length: Select the menu element 9117.4.6 Stop BitsSelect the number of stop bits.Stop bit 1: Select the menu element 92Stop bit 2: Select the menu element 9317.4.7 LimiterThe „limiter” is used to separate single data or commands from each other. Thelimiter is to be set in the following way:Setting to CR(0DH): Select the menu element 94Setting to LF(0AH): Select the menu element 95Setting to CR+LF(0D0AH): Select the menu element 96